Opens in a new tab

A Laterally-Implanted Piezoresistive Skin Friction Sensor

This paper presents the packaging, fabrication, and calibration of a piezoresistive skin-friction sensor for the direct measurement of wall shear stress. The floating-element structure integrates laterally-implanted piezoresistors into the tether sidewalls to form a fully active Wheatstone bridge for electromechanical transduction. Experimental characterization at a bias voltage of 1.5 V indicates a sensitivity of 4.24 µV/Pa , a noise floor of 11.4 mPa/rtHz at 1 kHz , a linear response up to the maximum testing range of 2 Pa , and a flat dynamic response up to the testing limit of 6.7 kHz.

Hilton Head Workshop Jun 2008

Y. Li(1), V. Chandrasekharan(1),, B. Bertolucci(1), T. Nishida(1), L. Cattafesta(1), D.P. Arnold(1) and M. Sheplak(1)

Interdisciplinary Microsystems Group, University of Florida (1)