Opens in a new tab

A MEMS Microphone for Aeroacoustics Measurements

This paper presents the development of a silicon- micromachined, piezoresistive microphone for aeroacoustics measurements. The microphone consists of four dielectrically-isolated, single-crystal silicon piezoresistors on top of a 1500 A-thick, 210 um-diameter silicon-nitride membrane. The microphone was de- signed by employing a fundamental structural model coupled with a lumped-element energy model. Preliminary measurements indicate a sensitivity of 1µV/V/Pa and a linear response up to 155 dB. The measured and predicted sensitivity agree to within 3 dB.

AIAA Jan 1999

Mark Sheplak(1), John Seiner(2), Kenneth Breuer(3) and Martin Schmidt(3)

University of Florida (1), The University of Mississippi (2), Massachusetts Institute of Technology(3)