Opens in a new tab

Characterization of a Hydraulically Smooth Wall Shear Stress Sensor for Low-Speed Wind Tunnel Applications

This paper describes the development and experimental characterization of a wall shear stress sensor system designed to be fully non-invasive to low-speed aerodynamic flows. The differential capacitive microelectromechanical systems (MEMS) sensor is fabricated with backside electrical contacts, eliminating the need for front side wire bonds. Packaged in a modular housing with interchangeable components, the result is a device with less than +/-40 um of deviation over the wetted surface, satisfying the requirements of hydraulic smoothness for low speeds less than 30 m/s. Mean shear stress calibration includes a sensitivity of 38.8 mV/Pa with a +/-0.2% full-scale accuracy for a 50 Pa maximum wall shear stress. Dynamic calibration includes a 3 dB bandwidth of 1.4 kHz with a 45.0 mV/Pa sensitivity and 60 uPa minimum detectable shear stress at 1.128 kHz.

AIAA Jan 2017

David Mills(1), Casey Barnard(2), Mark Sheplak(2)

Interdisciplinary Consulting Corporation (1), Interdisciplinary Microsystems Group, University of Florida (2)